  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 92
H-Class 83
Advanced Entry Calibration
Step
Action Displayed Message Comment
A Turn ON the printer.
CANNOT CALIBRATE
-or-
UNCALIBRATED
Wait briefly for
initialization to complete;
see Section 4.1.1.
B
Press ENTER to access
MEDIA SETTINGS. Then
using the UP Button, scroll
to SENSOR CALIBRATION.
MEDIA SETTINGS
ADVANCED MENU must
be enabled to access the
calibration selections.
C
Press ENTER to access
SENSOR CALIBRATION.
Using the DOWN Button,
highlight ADVANCED
ENTRY then press ENTER.
ADVANCED ENTRY
Press EXIT to abort this
procedure.
D
Scroll to TRAN SENSOR
GAIN (or REFL SENSOR
GAIN, if using reflective
media) then press ENTER.
TRAN SENSOR GAIN
yyy (0 - 31)
25
The following examples
detail die-cut media
calibration; however,
unless otherwise noted,
the reflective media
procedure is the same.
E
Install media. Position the
Media Sensor under the
label, then lower and latch
the Printhead Assembly.
TRAN SENSOR GAIN
yyy (0 - 31)
25
See Section 3.2 for
sensor adjustment
instructions.
Do not position the
Media Sensor under a
perforation; and if
using preprinted
media, ensure the
label area placed over
the sensor is free of
text, graphics, lines,
etc.